Hg5® Troubleshooting and Maintenance
The Hg5® is designed to provide years of trouble-free service, with minimal attention. In the unlikely
event of system related problems, please consult the troubleshooting and maintenance chart in your
Hg5 manual, or call SolmeteX® for assistance.
Problem: Low vacuum at the hand piece
If the vacuum gauge reads normally but there is little or no vacuum to the hand piece, there is
probably a clog or a leak between the hand piece and the Hg5®.
If the vacuum gauge reads lower than normal these are the possible causes:
Did you just replace the container?
Yes:
- Check installation of the new container
- If the vacuum is low there is a possibility that the o-rings on the container did not seal into the
receivers.
- Remove the container, check the o-rings and re-install per the installation procedure.
No:
- Check solid collection cup(s) - Clean or replace element/screen if necessary.
- Check operation of the vacuum pump.
- Listen for vacuum leaks.
- Check all connections for breaks or cracks.
- Check all flexible hose for kinks, breaks, or loose clamps.
Problem: No vacuum or liquid flow in the system
- Check solid collection cup(s) - Clean or replace element/screen if necessary.
If the problem cannot be solved easily, e-mail us for support.
?
?
?
?
/